Colorimetry: Principles to Evaluation

LEARNING AIM B - Colorimetry in Practice

1 / 58
volgende
Slide 1: Slide
newEditor

In deze les zitten 58 slides, met interactieve quizzen en tekstslides.

Dit is niet oké

Onderdelen in deze les

LEARNING AIM B - Colorimetry in Practice

This lesson sequence follows the full practical workflow: principles of light absorption, calibration, preparation of standards, construction of a calibration curve, determination of an unknown, processing results, and evaluation of data quality.

LEARNING AIM B

Use this to surface prior ideas such as light, colour, absorbance, transmission, concentration, filter, wavelength and cuvette. Pick up misconceptions such as 'it measures colour directly' for later correction.

Learning goal - BP3, BM3, BD3++

Explain how colorimetry works

Use a calibrated colorimeter to obtain results

Interpret results to determine and evaluate concentration.

Deze slide heeft geen instructies

LEARNING OBJECTIVES

  1. Operate a colorimeter safely and accurately, identifying its main components and calibrating it using an appropriate blank.

  2. Prepare standard solutions and serial dilutions using appropriate laboratory equipment and accurate volume measurements.

  3. Identify independent, dependent and controlled variables in a colorimetry investigation.

  4. Collect and record absorbance data for standard solutions and an unknown sample in a correctly organised results table.

  5. Construct and interpret a calibration curve by plotting absorbance against concentration and using it to estimate the concentration of an unknown sample.

  6. Process and analyse quantitative data, calculating means, ranges and percentage differences where appropriate, identifying anomalous results and explaining the importance of repeat measurements.

  7. Interpret and evaluate experimental results, distinguishing between accuracy, precision, reliability and validity, and suggesting evidence-based improvements to the method.

Deze slide heeft geen instructies

Light passing through a solution

  • A colorimeter sends light through a liquid sample

  • Some light is absorbed by the solution

  • The remaining light is transmitted to the detector

  • The instrument compares transmitted light with the original light

Link to students' likely prediction: darker or more concentrated solutions usually let less light through.

Absorption and transmission

Absorption and transmission are linked. If a solution absorbs more light, less light reaches the detector. In simple terms, high absorbance means low transmission, while low absorbance means high transmission.

Avoid heavy maths here unless needed. The key idea is the inverse relationship between absorbance and transmitted light.

Why colour matters

  • A coloured solution does not absorb all wavelengths equally

  • It absorbs some wavelengths strongly and transmits others more easily

  • The colour we see is linked to the light that is transmitted

Students often think the solution absorbs the same amount of every colour of light. Emphasise selective absorption.

Choosing the wavelength

  • The wavelength or filter must match the sample being tested

  • The best choice is the wavelength the solution absorbs most strongly

  • This gives a clearer difference between concentrations

  • Using the same wavelength keeps readings comparable

Bring in complementary colours here: for example, a blue solution is measured using light in the complementary region because that is absorbed more strongly.

Complementary Colours Matter

  • A coloured solution appears the colour it transmits most strongly, so it tends to absorb the complementary colour. For example, a blue solution is measured best using an orange-red region of light. Choosing a complementary colour filter helps produce a clearer absorbance difference between samples.

Deze slide heeft geen instructies

Absorbance and concentration

  • More particles in solution can absorb more light

  • As concentration increases, absorbance usually increases

  • As concentration increases, transmission usually decreases

  • This relationship allows concentration to be estimated from absorbance

State that this is most useful within the instrument's linear working range, which can be revisited later when evaluating data.

Which change usually happens when a solution absorbs more light?

A

Lower transmission

B

Higher transmission

C

Lower concentration

D

Higher wavelength

Deze slide heeft geen instructies

Which change usually happens when a solution absorbs more light?

A

Lower transmission

B

Higher transmission

C

Lower concentration

D

Higher wavelength

Checks the inverse link between absorbance and transmission from slides 2 to 6. Distractors reflect common confusion between transmission, concentration and wavelength.

Fill in the blanks

A colorimeter measures how much light passes through a sample.

Some light is xxxxxx by the solution and the rest is xxxxxxxxx to the detector. The best wavelength is the one the solution absorbs most xxxxxxxx. As concentration increases, xxxxxxxxxx usually increases.

Wordbank - absorbance, absorbed, strongly, transmitted

Deze slide heeft geen instructies

A colorimeter measures how much light passes through a sample.

Some light is absorbed by the solution and the rest is transmitted to the detector.
The best wavelength is the one the solution absorbs most strongly.
As concentration increases, absorbance usually increases.

Deze slide heeft geen instructies

Why must the same wavelength be used for all compared samples?

Using the same wavelength ensures that absorbance differences arise only from concentration differences, preserving the validity of the Beer–Lambert relationship and allowing accurate comparison across samples.


You need to expand upon this point





Deze slide heeft geen instructies

Main parts of a colorimeter

  • Light source

  • Wavelength selector or filter

  • Cuvette holder

  • Detector

  • Display or reading output

If using a diagram, point out that model designs vary, but these functional parts are common.

What calibration does

Calibration sets the instrument to a known starting point before measuring samples. This makes later absorbance readings meaningful, because the colorimeter has been adjusted using a blank rather than an unknown solution.

Stress that calibration is not optional preparation; it directly affects the trustworthiness of every later reading.

Using an appropriate blank

  • The blank contains everything except the coloured solute

  • It is often distilled water or the solvent used

  • The blank accounts for light loss from the cuvette and solvent

  • Zeroing with a blank makes the sample reading more accurate

If learners ask why not use an empty cuvette, explain that the solvent and cuvette also affect light transmission.

Calibration sequence

Switch on the colorimeter → select the correct wavelength or filter → prepare the blank in a clean cuvette → place the blank correctly in the holder → set the instrument to zero or follow its calibration command.

Exact button labels vary by instrument, so connect this general sequence to the local model in the lab.

Handling cuvettes correctly

  • Use clean, undamaged cuvettes

  • Remove fingerprints and external droplets

  • Fill to a consistent level

  • Keep the clear sides in the light path

  • Use the same orientation each time

Explain that scratches, smears and changing orientation can alter the amount of light reaching the detector.

If calibration is poor

  • The zero point is wrong

  • Absorbance values can be shifted up or down

  • Comparisons between samples become less reliable

  • Any calibration curve built from these readings may be misleading

This slide sets up later evaluation work: early procedural errors can propagate through the whole dataset.

What is the main purpose of a blank during calibration?

A

To darken the sample

B

To clean the detector

C

To set a baseline

D

To increase absorbance

Deze slide heeft geen instructies

What is the main purpose of a blank during calibration?

A

To darken the sample

B

To clean the detector

C

To set a baseline

D

To increase absorbance

Targets slides 8 to 12. Students should recognise that the blank accounts for the solvent and cuvette, creating a known starting point.

Can you put these steps in the correct order?

  1. Record the absorbance value for each standard and repeat readings to check consistency.

  2. Select the correct wavelength or filter, usually the complementary colour to the solution.

  3. Insert the blank cuvette and press “Zero” until the display reads 0 absorbance or 100% transmittance.

  4. Plot absorbance on the y‑axis against concentration on the x‑axis to create a calibration curve.

  5. Switch on the colorimeter and allow it to warm up for 5–10 minutes.

  6. Prepare a series of known concentrations of the coloured solution and place each in a clean cuvette.

  7. Wipe the cuvette with lint‑free tissue to remove fingerprints and ensure clear sides face the light path.

  8. Insert each standard cuvette into the colorimeter and measure its absorbance.

  9. Check the calibration curve for linearity, outliers, and increasing absorbance with concentration.

  10. Fill a clean cuvette three‑quarters full with the blank (distilled water or solvent).

Deze slide heeft geen instructies

Answers

  1. Switch on the colorimeter and warm up.

  2. Select the correct filter.

  3. Prepare the blank cuvette.

  4. Wipe and align the cuvette.

  5. Zero the colorimeter with the blank.

  6. Prepare calibration standards.

  7. Insert each standard and measure absorbance.

  8. Record absorbance values and repeat if needed.

  9. Plot the calibration curve.

  10. Check calibration quality.

Deze slide heeft geen instructies

Standard solutions

  • A standard solution has a known concentration

  • Several standards are needed to build a calibration curve

  • The standards must cover a suitable concentration range

  • Their concentrations must be prepared accurately

Link back to purpose: without known concentrations, the colorimeter cannot be used quantitatively.

Extension Questions

  • Why must the blank be used before measuring standards?

  • What happens if fingerprints remain on the cuvette?

  • Why should absorbance increase with concentration?

  • What would you do if your calibration curve is not linear?

  • Why is the complementary colour filter chosen?

Deze slide heeft geen instructies

Accurate volume measurement

  • Use appropriate equipment for the volume required

  • Measure at eye level to reduce reading error

  • Transfer liquids carefully to avoid loss

  • Mix thoroughly so concentration is uniform

  • Label each solution clearly

You can refer to pipettes, measuring cylinders and volumetric flasks depending on what is available.

Serial dilution

A serial dilution is made by taking a measured volume from one solution and diluting it to make the next, lower concentration. This method is useful for producing a series of standards with known and regular concentration changes.

If needed, demonstrate with one simple example verbally rather than placing calculations on the slide.

Controlling variables

  • Keep the same total volume for each standard

  • Use the same solvent throughout

  • Use the same mixing method

  • Label and store solutions consistently

  • Control these variables so differences are due to concentration

Students often focus only on calculations. Emphasise that practical consistency is part of data quality.

Why dilution errors matter

  • An incorrect volume gives the wrong concentration

  • Wrong standards produce a flawed calibration curve

  • A flawed curve leads to poor estimates for unknown samples

  • Care in preparation affects every later result

Make the chain explicit from standards to final unknown concentration.

Why must standard solutions be prepared accurately?

A

To reduce light colour

B

To fix detector damage

C

To make labels easier

D

To build a valid curve

Connects standards, dilution accuracy and the quality of the calibration curve from slides 13 to 17.

Why must standard solutions be prepared accurately?

A

To reduce light colour

B

To fix detector damage

C

To make labels easier

D

To build a valid curve

Connects standards, dilution accuracy and the quality of the calibration curve from slides 13 to 17.

Measuring the standards

  • Use the calibrated colorimeter

  • Measure each standard at the same wavelength

  • Check cuvette cleanliness and orientation each time

  • Repeat readings where appropriate

  • Record absorbance carefully

Point out that consistency in handling is just as important as consistency in concentration.

Recording results clearly

  • Organise results in a table

  • Include concentration and absorbance columns

  • Add units for concentration

  • Keep repeat readings in separate columns

  • Leave a clear space for a mean if calculated

Good recording reduces transcription errors and makes graph plotting easier.

Plotting a calibration curve

  • Plot concentration on the x-axis

  • Plot absorbance on the y-axis

  • Mark points accurately

  • Draw a suitable trend line

  • Look for the overall pattern in the data

Students commonly reverse the axes; emphasise that concentration is the independent variable here.

Interpreting the trend

  • A higher concentration usually gives a higher absorbance

  • Many calibration curves are approximately linear over a useful range

  • A point far from the trend may be anomalous

  • Unexpected points should be checked against method and measurements

This is the place to introduce the idea of linearity without overloading students with theory.

Using the curve for an unknown

Measure the absorbance of the unknown sample at the same wavelength and under the same conditions as the standards. Its concentration can then be estimated by matching that absorbance to the calibration curve.

Stress that the unknown must be measured using the same procedure as the standards, otherwise the comparison is weaker.

Reading concentration from the graph

  • Find the absorbance value on the y-axis

  • Move across to the trend line

  • Move down to the x-axis

  • Read the estimated concentration

  • Record the value with appropriate units

If an equation is available from a spreadsheet trend line, it can also be used, but graph reading should remain the core method here.

Repeat measurements and anomalies

  • Repeats help check consistency

  • Similar repeat readings increase confidence

  • A very different reading may be anomalous

  • Anomalies should be investigated, not ignored without reason

You can mention possible causes such as bubbles, cuvette smears, poor mixing or incorrect zeroing.

Limits of graph estimates

  • Values read from a graph are estimates

  • Uncertainty increases if the line is unclear

  • Results outside the main trend range are less dependable

  • Precision depends on both the graph quality and the measurements

Good place to introduce the idea that interpolation is stronger than extrapolation, if appropriate for the group.

On a calibration curve, what goes on the y-axis?

A

Concentration

B

Time

C

Absorbance

D

Wavelength

Checks a common error from slides 18 to 25: concentration is the independent variable on the x-axis, absorbance on the y-axis.

On a calibration curve, what goes on the y-axis?

A

Concentration

B

Time

C

Absorbance

D

Wavelength

Checks a common error from slides 18 to 25: concentration is the independent variable on the x-axis, absorbance on the y-axis.

Processing colorimetry data

  • Raw readings can be summarised to show patterns more clearly

  • Common processing includes mean, range and percentage difference

  • Processed data should still stay linked to the original readings

Distinguish between raw data and processed data clearly before moving into calculations.

Mean, range and percentage difference

  • Mean: the average of repeated readings

  • Range: the difference between the highest and lowest reading

  • Percentage difference: the size of a difference relative to the values being compared

Keep the definitions practical and linked to repeated absorbance readings rather than abstract statistics.

Presenting processed data

  • Use clear tables with headings and units

  • Choose graphs that match the data

  • Keep axes labelled correctly

  • Make processed results easy to compare between samples or groups

If using spreadsheets, model formula use and careful checking of copied cells.

Recognising patterns in results

  • A consistent trend supports the expected relationship

  • Small ranges suggest readings are close together

  • Large differences may point to technique problems or anomalies

  • Interpretation should be based on the actual numbers

This prepares students for evidence-based evaluation rather than vague comments such as 'the results were good'.

What does a small range in repeat readings suggest?

A

Low precision

B

High precision

C

Poor validity

D

High percentage difference

Uses slides 26 to 29 to distinguish what processed data can show about consistency.

What does a small range in repeat readings suggest?

A

Low precision

B

High precision

C

Poor validity

D

High percentage difference

Uses slides 26 to 29 to distinguish what processed data can show about consistency.

Judging result quality

  • Accuracy: closeness to the true or accepted value

  • Precision: closeness of repeated readings to each other

  • Reliability: whether results are consistent when repeated

  • Validity: whether the method measures what it is meant to measure

Students often confuse precision with accuracy. Use a simple contrast if needed: tightly grouped wrong values are precise but not accurate.

Comparing results quantitatively

  • Compare means, ranges or percentage differences

  • Check whether groups show similar trends

  • Use numerical evidence to support conclusions

  • Avoid general statements without data

Model sentence stems orally if helpful, such as 'Group A had a lower mean absorbance than Group B by...'.

Evaluating the method

  • Consider calibration quality

  • Consider dilution accuracy

  • Consider cuvette handling

  • Consider instrument limits and graph uncertainty

  • Link each limitation to its likely effect on the final concentration

Push students to explain the effect, not just name the issue. For example, a dirty cuvette may falsely increase absorbance.

Evidence-based conclusions

A strong evaluation uses specific results to judge how trustworthy the final concentration is. The best conclusion links numerical evidence, procedural quality and recognised limitations in one clear account.

This slide resolves the lesson arc: colourimetry is not just taking a reading, but building confidence in what that reading means.

Which term means closeness to the true or accepted value?

A

Reliability

B

Validity

C

Accuracy

D

Precision

Checks the distinction among accuracy, precision, reliability and validity from slides 30 to 33.

Which term means closeness to the true or accepted value?

A

Reliability

B

Validity

C

Accuracy

D

Precision

Checks the distinction among accuracy, precision, reliability and validity from slides 30 to 33.

Which part of colorimetry now feels most secure to you?

1

Light and absorbance

2

Calibration

3

Dilutions

4

Using the graph

5

Evaluation

Deze slide heeft geen instructies

Glossary

absorbance - a measure of how much light a solution absorbs

transmission - the passage of light through a sample

wavelength - the distance between equivalent points on a wave, used here to describe the colour of light

filter - a component that selects a particular range of wavelengths

colorimeter - an instrument that measures how much light a solution absorbs

cuvette - a small transparent container used to hold a sample in a colorimeter

blank - a reference solution used to set the instrument baseline

Deze slide heeft geen instructies

Glossary

standard solution - a solution with a known concentration

serial dilution - stepwise dilution used to make a series of lower concentrations

calibration - setting an instrument to a known starting point before measuring samples

calibration curve - a graph showing the relationship between concentration and absorbance

anomalous result - a result that does not fit the pattern of the other data

precision - how close repeated results are to each other

validity - how well a method measures what it is intended to measure